To use this function you have to enable JavaScript and allow popup windows for this domain.
 
Analytical Columns
Pro Family : YMC-Pack Pro C18
 

Separation of extremely basic compounds
Tricyclic antidepressants
Column: 150 x 4.6 mm i.d.
Eluent: 20 mM potassium phosphate buffer (pH 6.9) / methanol (25/75)
Flow rate: 1.0 ml/min
Temperature: 37°C
Detection: UV at 254 nm

Separation of peptides under LC-MS compatible conditions
Tetracycline antibiotics
Column: YMC-Pack Pro C18 (5 μm, 12 nm, 150 x 4.6 mm)
Eluent: acetonitrile / 10mM oxalic acid (10/90)
Flow: 1.0 ml/min
Detection: UV at 280 nm, 0.13 AUFS
Temperature: 37 ºC
Injection: 6 μl (0.2 mg/ml)

Flavonoids
Antiepileptics
Column: YMC-Pack Pro C18 (5 μm, 12 nm, 150 x 4.6 mm)
Eluent: 50 mM NH4H2PO4-(NH4)2HPO4 (pH 6.7) / methanol (50/50)
Flow: 0.7 ml/min
Detection: UV at 215 nm, 0.32 AUFS
Temperature: 30 ºC
Injection: 8 μl (0.035  0.7 mg/ml)

Antiepileptics
Flavonoids
Column: YMC-Pack Pro C18 (5 μm, 12 nm, 150 x 4.6 mm)
Eluent: acetonitrile / 10 mM H3PO4 (35/65)
Flow: 1.0 ml/min
Detection: UV at 280 nm, 0.13 AUFS
Temperature: 37 ºC
Injection: 5 μl (0.05 mg/ml)

Separation of related cardiac glycosides
Cardiac glycosides
Column: YMC-Pack Pro C18 (5 μm, 12 nm, 75 x 4.6 mm)
Eluent: acetonitrile / water (10/90)
Flow: 1.0 ml/min
Detection: UV at 220 nm, 0.26 AUFS
Temperature: 37 ºC
Injection: 5 μl (0.2, 0.4 mg/ml)
For more applications please refer to our “Application Data Collections” or contact us directly.

Column care
YMC Pack Pro C18 is stable towards hydrolysis between pH 2-8. Remove acid and buffer salts before storage. Store the column in methanol / water = 70/30.
For detailed information please refer to the “Column Care and Use Instructions” which are shipped with each analytical column.
 


© 2006 -2012 YMC Europe GMBH

 